Isolation transformers

REOMED II 1120

Unique Selling Point


  • - IEC 60601-1 (ed 3.1)

    - IEC 60601-1-2 (ed 4.0)

    - Certified according to NRTL

  • built-in circuit breakers 
  • Good heat dissipation
  • Compact dimensions
  • Solid aluminum casing
  • Green illuminated mains switch
  • Low total weight
  • Prepared for IEC60601-1:2005 (3rd. Edition)
  • High efficiency with a very low magnetic field
  • Low noise
  • Easy mounting
  • Very low non-load current
  • Low voltage variation
  • Compact design
  • Connection (PTC) for external temperature monitoring

Description


The REOMED II isolation transformer provides reliable leakage current reduction in medical applications. Safe isolation on the input side is affected by using high quality materials and first class production methods. The isolation transformers are designed for low internal losses and so achieve very low no-load losses.
Overload and short-circuit protection on the input and output side is affected by an integral circuit breaker.

Application area:

  • Medical systems

Technical Data


  • Rated power : 1120 VA
  • Input voltage : 230 V
Your contact: +1 (317) 899-1395 • info@reo-usa.com

REOMED II 1120

Isolation transformers

REOMED II

1120

Nominal input power [VA]

1,120

Input voltage [V]

230

Output voltage [V]

230

Mains frequency [Hz]

50/60

Input power plug

1 piece IE C/EN 60320 (C20) UL 498, CSA C22.2 no. 42

Output power plug

6-piece IEC/EN60320-2-2 (C13) UL 498 CSA C22.2 no. 42

Power switch [0-1]

yes

Circuit breaker [A]

6

Sec. circuit breaker [A]

5

Thermal overload protection

yes

Temperature sensor

yes, in connection with insulation monitor

Start-up current limiter

NTC (option 10) or electronic (option 50)

Max. output current [A]

4.7

Earth leakage current 254V [µA]         

<500

Insulation / protection resistance [Ω]

>2M/<0.1

Test voltage

primary-secondary.: 4kV AC; primary housing / secondary housing: 1.5kV AC

Protection class/med. class/Type

I/I

Protection IP

IP20

Ambient temperature [°C]

±0 - +40

Ambient air humidity [%]

Relative 30 - 75

Max. altitude [m]

2,000

Equi-potential earthing (DIN42801)

POAG - S6/15

Housing, Surface

Powder-coated metal - RAL 7035

Dimensions (BxHxT) [mm]

235 x 115* x 325 (with rubber foot 122)

Weight [kg]

13.0

Approval / principles  /
Conformity CE

IEC60601-1:2005/AMD1:2012; ANSI/AAMIES 60601-1:2005/CR/2012

CAN/CSA C22.2 No. 60601-1:14 (Medical) USA and Canada

IEC60601–1-2:2014(Partly); EN/IEC 61000-3-2:2014; EN/IEC 61000-3-3:2013

Accessories (Options)

Power line

Country-specific

Trigger safety

Mounting kit (1x trigger guard, 1x screws kit)

Mounting rails

Mounting kit (2x mounting rails, 4x hexagon screw)

Your contact: +1 (317) 899-1395 • info@reo-usa.com

REOMED II 1120

Isolation transformers


Type

H (mm)

B (mm)

T (mm)

Weight (kg)

REOMED II 1120

115 (122*)

235

325

13.0

*with rubber foot