Download PDF
Your contact: +1 (317) 899-1395 • info@reo-usa.com

REOMED 300

Isolation Transformer for the Medical Technology

REOMED 300

Unique Selling Point


  • IEC 60601-1 (ed. 3.1)
  • IEC 60601-1-2 (ed. 4.0)
  • Certificated according to NRTL
  • Optional shield winding between primary and secondary current
  • Variants with filter and surge protection
  • High efficiency with a very low magnetic field
  • Wide range of options
  • Small weight
  • Short-circuit and overload protection
  • Integral start-up current limiter
  • Solid aluminum casing
  • Equipotential bonding pin DIN 42801
  • Output sockets EN 60320

Description


The TÜV-certified REOMED isolating transformers are proven and reliable equipment for use with all electrical systems in a medical environment. They reduce the leakage current and thus help to ensure the safety of patients.
REOMED isolating transformers are characterized by their very low magnetic stray-fields and reliability, whilst also providing high efficiency and easy connectivity.

Application area:

  • Medical systems

Technical Data


  • Rated power : 300 VA
  • Input voltage : 115 - 230 V

REOMED Maßbild

Type

H (mm)

B (mm)

T (mm)

Weight (kg)

REOMED 300

85 (90*)

150

240

4.5

*with rubber feet

Your contact: +1 (317) 899-1395 • info@reo-usa.com

REOMED 300

Isolation Transformer for the Medical Technology

REOMED

300

Nominal input power [VA]

300

300

300

300

300

300

Input voltage [V]

115/230

230

115

115

230

115/230

Output voltage [V]

230

230

230

115

115

115/230

Mains frequency [Hz]

50/60

50/60

50/60

50/60

50/60

50/60

Input power plug

1 piece IE C/EN 60320 (C14) UL 498, CSA C22.2 no. 42

Output power plug

4 piece IEC/EN60320-2-2 (C13) UL 498 CSA C22.2 no. 42

Power switch [0-1]

yes

yes

yes

yes

yes

yes

Input fuse [A]

T3, 15
T1, 60

T1, 60

T3, 15

T3, 15

T1, 60

T3, 15
T1, 60

Thermal overload protection [°C]

120

120

120

120

120

120

Max. output current [A]

1.30

1.30

1.30

2.60

2.60

2.60/1.30

Output fuse [A]

-

-

-

-

-

-

Start-up current limiter

yes

yes

yes

yes

yes

yes

Standard

NTC

NTC

NTC

NTC

NTC

NTC

Earth leakage current 127/254V [µA]         

<300/500

<300/500

<300/500

<300/500

<300/500

<300/500

Insulation / protection resistance [Ω]

>2M/<0.1

>2M/<0.1

>2M/<0.1

>2M/<0.1

>2M/<0.1

>2M/<0.1

Test voltage

primary-secondary: 4kV AC; primary housing / secondary housing: 1.5kV AC

Protection class/med. class/Type

I/I

I/I

I/I

I/I

I/I

I/I

Protection IP

IP20

IP20

IP20

IP20

IP20

IP20

Ambient temperature [°C]

0...40

0...40

0...40

0...40

0...40

0...40

Ambient Air Humidity [%]

Relative
30 - 75

Relative
30 - 75

Relative
30 - 75

Relative
30 - 75

Relative
30 - 75

Relative
30 - 75

Max. altitude [m/feet]

3000/9864

3000/9864

3000/9864

3000/9864

3000/9864

3000/9864

Equi-potential earthing (DIN42801)

POAG - S6/15

Housing, Surface

Powder-coated metal - RAL 7035

Dimensions (BxHxT) [mm/in]

150x85*x240 (*with rubber feet 90) / 5.9x3.34*x9.54 (*with rubber feet 3.54)

Weight [kg/lbs]

4.5/10

Approval / principles  /
Conformity CE

IEC60601-1:2005/AMD1:2012; ANSI/AAMIES 60601-1:2005/CR/2012 CAN/CSA C22.2 No. 60601-1:14 (Medical) USA and Canada IEC60601–1-2:2014(Partly); EN/IEC 61000-3-2:2014; EN/IEC 61000-3-3:2013

Accessories (Options)

Power line

Country-specific

Pull-off protection

Mechanic retaining kit

Mounting rails

Mounting kit (2x mounting rails, 4x hexagon screw)