The TÜV-certified REOMED isolating transformers are proven and reliable equipment for use with all electrical systems in a medical environment. They reduce the leakage current and thus help to ensure the safety of patients.
REOMED isolating transformers are characterized by their very low magnetic stray-fields and reliability, whilst also providing high efficiency and easy connectivity.
Application area:
Type |
H (mm) |
B (mm) |
T (mm) |
Weight (kg) |
REOMED 300W |
95 (100*) |
170 |
270 |
7.5 |
*with rubber feet
REOMED |
300W |
Nominal input power [VA] |
300 |
Input voltage [V] |
100-130/200-250 |
Output voltage [V] |
100-130/200-250 |
Mains frequency [Hz] |
50/60 |
Input power plug |
1 piece IE C/EN 60320 (C14) UL 498, CSA C22.2 no. 42 |
Output power plug |
6 piece IEC/EN60320-2-2 (C13) UL 498 CSA C22.2 no. 42 |
Power switch [0-1] |
yes |
Input fuse [A] |
T3, 15 / T1, 60 |
Thermal overload protection [°C] |
120 |
Max. output current [A] |
3.00-2.30 / 1.50-1.20 |
Output fuse [A] |
- |
Start-up current limiter |
yes |
Standard |
electronic |
Earth leakage current 127/254V [µA] |
<300/500 |
Insulation / protection resistance [Ω] |
>2M/<0.1 |
Test voltage |
primary-secondary: 4kV AC; primary hosuing / secondary housing: 1.5kV AC |
Protection class/med. class/Type |
I/I |
Protection IP |
IP20 |
Ambient temperature [°C] |
0 ... 40 |
Ambient air humidity [%] |
Relative 30 - 75 |
Max. altitude [m/feet] |
3000/9864 |
Equi-potential earthing (DIN42801) |
POAG - S6/15 |
Housing, Surface |
Powder-coated metal - RAL 7035 |
Dimensions (BxHxT) [mm/in] |
170x95*x270 (*with rubber feet 100)/6.70x3.34*x10.63 (*with rubber feet 3.54) |
Weight [kg/lbs] |
7.5/16.5 |
Approval / principles / |
IEC60601-1:2005/AMD1:2012; ANSI/AAMIES 60601-1:2005/CR/2012 |
Accessories (Options) |
|
Power line |
Country-specific |
Pull-off protection |
Mechanic retaining kit |
Mounting rails |
Mounting kit (2x mounting rails, 4x hexagon screw) |