Download PDF
Your contact: +1 (317) 899-1395 • info@reo-usa.com

REOMED 600

Isolation Transformers

REOMED 600

Unique Selling Point


  • IEC 60601-1 (ed. 3.1)
  • IEC 60601-1-2 (ed. 4.0)
  • Certificated according to NRTL
  • Optional shield winding between primary and secondary current
  • Variants with filter and surge protection
  • High efficiency with a very low magnetic field
  • Wide range of options
  • Small weight
  • Short-circuit and overload protection
  • Integral start-up current limiter
  • Solid aluminum casing
  • Equipotential bonding pin DIN 42801
  • Output sockets EN 60320

Description


The TÜV-certified REOMED isolating trans-ormers are proven and reliable equipment for use with all electrical systems in a medical environment. They reduce the leakage current and thus help to ensure the safety of patients.
REOMED isolating transformers are characterized by their very low magnetic stray-fields and reliability, whilst also providing high efficiency and easy connectivity.

Application area:

  • Medical systems

Technical Data


  • Rated power : 600 VA
  • Input voltage : 115 - 230 V

Type

H (mm)

B (mm)

T (mm)

Weight (kg)

REOMED 600

95 (100*)

170

270

7.5

*with rubber feet

Your contact: +1 (317) 899-1395 • info@reo-usa.com

REOMED 600

Isolation Transformers

REOMED

600

Nominal input power [VA]

600

600

600

600

600

600

Input voltage [V]

115/230

230

115

115

230

115/230

Output voltage [V]

230

230

230

115

115

115/230

Mains frequency [Hz]

50/60

50/60

50/60

50/60

50/60

50/60

Input power plug

1 piece IE C/EN 60320 (C14) UL 498, CSA C22.2 no. 42

Output power plug

6-piece IEC/EN60320-2-2 (C13) UL 498 CSA C22.2 no. 42

Power switch [0-1]

yes

yes

yes

yes

yes

yes

Input fuse [A]

T6.3/T3.15

T3, 15

T6, 3

T6, 3

T3, 15

T6.3/T3.15

Thermal overload protection [°C]

120

120

120

120

120

120

Max. output current [A]

2.60

2.60

2.60

5.20

5.20

5.20/2.60

Output fuse [A]

-

-

-

-

-

-

Start-up current limiter

yes

yes

yes

yes

yes

yes

Standard

NTC

NTC

NTC

NTC

electronic

electronic

Earth leakage current 127/254V [µA]         

<300/500

<300/500

<300/500

<300/500

<300/500

<300/500

Insulation / protection resistance [Ω]

>2M/<0.1

>2M/<0.1

>2M/<0.1

>2M/<0.1

>2M/<0.1

>2M/<0.1

Test voltage

primary-secondary.: 4kV AC; primary housing / secondary housing: 1.5kV AC

Protection class/med. class/Type

I/I

I/I

I/I

I/I

I/I

I/I

Protection IP

IP20

IP20

IP20

IP20

IP20

IP20

Ambient temperature [°C]

0...40

0...40

0...40

0...40

0...40

0...40

Ambient air humidity [%]

Relative
30 - 75

Relative
30 - 75

Relative
30 - 75

Relative
30 - 75

Relative
30 - 75

Relative
30 - 75

 

           

Max. altitude [m/feet]

3,000/9,864

3,000/9,864

3,000/9,864

3,000/9,864

3,000/9,864

3,000/9,864

Equi-potential earthing (DIN42801)

POAG - S6/15

Housing, Surface

Powder-coated metal - RAL 7035

Dimensions (BxHxT) [mm/in]

170x95*x270 (*with rubber feet 100)/6.70x3.34*x10.63 (*with rubber feet 3.54)

Weight [kg/lbs]

7.5/16.5

Approval / principles /
Conformity CE

IEC60601-1:2005/AMD1:2012; ANSI/AAMIES 60601-1:2005/CR/2012 CAN/CSA C22.2 No. 60601-1:14 (Medical) USA and Canada IEC60601–1-2:2014(Partly); EN/IEC 61000-3-2:2014; EN/IEC 61000-3-3:2013

Accessories (Options)

Power line

Country-specific

Pull-off protection

Mechanic retaining kit

Mounting rails

Mounting kit (2x mounting rails, 4x hexagon screw)